Method & Thresholds
AlgorithmDeterministic rule-based (M1ADV v0.4.5-jms)
Core referenceGAT − (CCT − 540) / 25 — not a true IOP estimate
ORA measurementIOPcc (primary); CH, CRF, WFS as biomechanical context only
Corvis measurementbIOP (primary); SSI and SP-A1 as biomechanical context only
Both mode|ORA IOPcc − Corvis bIOP| reported; never fused; no primary device
CH noteOut-of-range flag when CH < 4 or > 18 mmHg (not a validated cutoff)
Quality mappingCorvis map provisional (corvis-map-0.1); raw label preserved
Refractive surgeryPrior LASIK / PRK / SMILE / RK → Core estimate marked “not applicable”
Session validationClinician confirmation required; encounter ID alone is not sufficient
Target-IOPNot eligible — result is contextual support, not a corrected IOP
Limitations:
CH/SSI/SP-A1 never modify the pressure value. Cross-device comparison requires same-session confirmation. Core estimate not suitable for target-IOP decisions. Not a certified medical device.
Disclaimer

Evidence-informed biomechanical IOP context based on ORA and/or Corvis ST device measurements. Does not replace clinical judgment or comprehensive tonometric evaluation.

This tool applies deterministic rule-based logic to contextualize intraocular pressure against corneal biomechanical parameters derived from the Ocular Response Analyser (ORA) and/or the Corvis ST. It is intended as a structured decision-support aid, not as a standalone diagnostic conclusion. The Core reference (GAT–CCT correction) is not a true IOP estimate and is not suitable for target-IOP comparisons.

Device measurements may be affected by corneal disease, ectasia, prior refractive surgery, media opacity, and instrument variability. Corneal Hysteresis, SSI, and SP-A1 are reported as biomechanical context only and never modify the pressure value. Cross-device comparison (ORA vs Corvis) requires same-session readings and clinician confirmation; without confirmation the comparison is flagged as unvalidated and reported as numerical only. Device pressures are never fused into a single value and no device is designated as primary.

Device-derived values are interpreted as biomechanical context and are never fused into a single pressure value. The M1 Core GAT/CCT reference is a contextual estimate, not true IOP. Target-IOP integration is not active in this module.

When only ORA or only Corvis is selected, biomechanical context is necessarily branch-specific. Cross-device comparison is available only in Both mode and requires same-session clinician confirmation.

The clinician retains full responsibility for all diagnostic and therapeutic decisions. This tool is intended as a decision-support aid, consistent with current glaucoma guidelines (EGS 2025), and is not a substitute for clinical evaluation. Not intended for patient self-use. Not a certified medical device.

Sources (Summary)
  • Ehlers N, Bramsen T, Sperling S. Applanation tonometry and central corneal thickness. Acta Ophthalmol. 1975;53(1):34–43.
  • Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg. 2005;31(1):156–162.
  • Medeiros FA, Meira-Freitas D, Lisboa R, Kuang TM, Zangwill LM, Weinreb RN. Corneal hysteresis as a risk factor for glaucoma progression: a prospective longitudinal study. Ophthalmology. 2013;120(8):1533–1540.
  • Joda AA, Shervin MM, Kook D, Elsheikh A. Development and validation of a correction equation for Corvis tonometry. Comput Methods Biomech Biomed Engin. 2016;19(9):943–953.
  • Eliasy A, Chen KJ, Vinciguerra R, Lopes BT, Abass A, Vinciguerra P, Ambrósio R Jr, Roberts CJ, Elsheikh A. Determination of corneal biomechanical behavior in-vivo for healthy eyes using CorVis ST tonometry: stress-strain index. Front Bioeng Biotechnol. 2019;7:105.
  • Vinciguerra R, Ambrósio R Jr, Elsheikh A, Roberts CJ, Lopes B, Morenghi E, Azzolini C, Vinciguerra P. Detection of keratoconus with a new biomechanical index. J Refract Surg. 2016;32(12):803–810.
  • European Glaucoma Society. Terminology and guidelines for glaucoma, 6th edition. Br J Ophthalmol. 2025;109(Suppl 1):1–212.
Version History
  • v0.4.7Jun 2026
    • Clinician-facing UI restyle: DM Sans / DM Mono fonts (matching OFTY family), card layout, M1 Core visual parity
    • ORA source result: single control (Single reading / Device-averaged output) replaces checkbox + enum pair
    • Session attestation (device-vs-GAT comparison validation) surfaced in main flow
    • Device metadata & timestamps merged into single optional collapsible (audit trail only)
    • Corneal modifiers: single-column layout, Prior refractive surgery select with explicit LASIK / PRK / SMILE / RK options
    • Info tabs reordered to match M3: Method & Thresholds → Disclaimer → Sources → Version History
    • Disclaimer: preview + “Read full disclaimer” expand pattern (M3-style)
    • Method & Thresholds: key–value table layout (M3-style meta-row)
    • Frozen engine bundle unchanged (9f9a0fe4…); 28/28 acceptance tests preserved
  • v0.4.5-jmsMay 2026
    • JMS verification release (frozen 50-file inventory, signed)
    • Engine bundle, schema, build manifest frozen

Engine algorithm — · Schema —